国产一区二区性情一级片,中文日本在线观看中文幕,年轻的母亲在线观看韩国,三级片视频一区

客戶咨詢熱線:
Technical articles技術文章
首頁 > 技術文章 >USP <787> Subvisible Particulate Matter in Therapeutic Protein Injections

USP <787> Subvisible Particulate Matter in Therapeutic Protein Injections

 更新時間:2015-05-15  點擊量:7112
  

Follow the same standardization steps as described in USP <788>

標準制定遵從USP <788>制定步驟

• Volume accuracy 

體積
• Flow rate

流速
• Sensor resolution

傳感器靈敏度
• Calibration

校準
• Count accuracy

計數

Two system controls tests are recommended

兩種系統控制測試

• Blank test: measure five aliquots of degassed particle-free water, the count must be less than 1 particle/mL > 10 µm.

測試:檢測脫氧純水,> 10 µm的粒子數不得超過1 個/mL
• System suitability verification; analyze the USP particle count reference standard and pass the count limits

系統適應性驗證:檢測分析USP標準粒子并低于限定值

Sample preparation

樣品準備

• If there is enough volume test individual units

確認每個樣品是否有足夠的體積
• If the volume is too small, mix units and combine the contents to obtain the required volume (typically 0.2-5.0 mL)

如果體積太小,將樣品混合直到符合測量的體積(典型0.2-5.0 mL)
• Degas the sample and gently mix again

輕輕上下晃動混合均勻

Analysis

• Analyze four aliquots

測量4次
• Count the particles in the size range of interest, including >10 & 25 µm

測量計數規定范圍內的粒子,包括>10 跟 >25 µm。
• Disregard the first result and average the second two

*次不計,求得后面兩次的平均值

Pass/Fail criteria

If the container volume < 100 mL;  less than 6000 particles/container > 10 µm and 600 particles/container > 25 µm

通過/不通過的標準范圍

如果標示裝量< 100 mL;> 10 µm 不超過6000個/瓶,> 25 µm 不超過600個/瓶

If the container volume > 100 mL; less than 25 particles/mL > 10 µm and 3 particles/mL > 25 µm. Also not exceed the per container limits for the < 100 mL  criteria above.

如果標示裝量> 100 mL,>10 µm 不超過25個/ml,> 25 µm 不超過10個/ml

The AccuSizer FX Nano system is ideal system for performing the USP <787> test for the following reasons

AccuSizer FX Nano是測量USP <787>的*選擇。

• Passes all standardization and system suitability tests

通過所有標準粒子與系統適應性測試
• Sensitivity down to 0.15 µm

靈敏度可以達到0.15 µm
• Higher concentration limit

可檢測高濃度樣品
• Accurate sample volume down to 250 µL

小測量體積為250 µL

 

Aggregated IgG protein before and after filtration as measured on the AccuSizer FX Nano

AccuSizer FX Nano關于聚合免疫蛋蛋白經過濾和未過濾的檢測譜圖

18无码粉嫩小泬无套在线观看 | 久久久一区二区三区不卡| 877无码视频在线观看| 人妻精品久久久一区二区| 樱桃视频影院在线播放| 日韩无码高清| 国产精品一区二区四av| 亚洲小说图片区综合在线| 天天躁日日躁狠狠躁免费麻豆| 一区二区 中文字幕 在线| 国产村长av一区二区三区| 欧美午夜性刺激在线播放| 天天色天天看天天综合网| 国产在线 | 中文| 色综合亚洲精品激情狠狠| 一区二区三区亚洲av电影| 国产精品va在线观看不| 综合成人网友亚洲偷自拍| 国产在线视频不卡一香蕉| 欧美日本一区二区三区久久| 日韩国产精品区一区二区| 国产精品毛片一区二区排| 精品无码一级毛片免费冫| 日本AⅤ精品无码黄皮片| 欧美亚日韩在线日本在线| 国产欧美精品一区一2区| 少妇性荡欲视频午夜免费| 亚洲无码三级片浪潮AV| 日韩一区日韩二区日韩三区| 精品一国产精品| 青青草国产精品欧美成人| 欧洲亚洲国产精华液| 久久久精品国产亚洲av麻| 91式老师啪在线国自产| 国产日本精品一区二区三区| 国产精品久久久久jk制服| 亚洲色自偷自拍另类小说| 亚洲成网站在线在线播放| 曰韩无码二三区中文字幕| 亚洲欧美国产另类91综合| 亚洲国产40页第11页|